GENERAL OPERATIONAL SPECIFICATIONS
1. The Monitron II Model # F00007A Wave Form Analyzer is based upon Digital PC computer logic,
employing internal conversion of standard line voltages either 115/50 or 220/50.
2. The standard Monitron Housings are designed to Interlock with the VDR4 Ventilators, providing
for easy straight on viewing.
3. The Montiron II Computer Screen provides for both Wave Form Presentations and combined
Digital functional read outs.
4. While the Monitron® II has no influence over the Operation of the VDR-4 Ventilator, a special
interface Harness interconnects the Monitron II to the Accessory Monitoring Ports of the VDR®-4
Servolator® Ventilator. (A failure of the Montron II has no influence upon the independent Ventilator
it is connected to.)
5. The Montiron II has three (3) selectable Proximal Airway Monitoring Pressure Scales, they are:
30, 60 and 120 cm H20.
6. The Monitron II has four (4) Sweep Speed Scales, 1, 2, 4 and 8 seconds, for full horizontal page
updating.
7. The Monitron independently reads the "CONVECTIVE" rate of Volumetric Tidal Exchange and
the "PERCUSSIVE" Sub Tidal Volume Delivery Rate, as well as:
a) I/E and i/e Ratios.
b) Convective Inspiratory and Expiatory Interval time in Seconds.
c) Mean and Peak Pressure Rise in cm H20.
d) PEEP and / or CPAP in cm H20.
e) PRIMARY WAVE FORMAT and / or DIGITAL DISPLAY mode selection.
8. The Monitron has an accessory RS232 Serial Printing Port, VGA port for connection to
a larger monitor, as well as Screen Freeze options.
9. The Monitron II is installationally interchangeable with the Monitron I with no influence over
Ventilator functions. |