1. Conventional Operational Source Pressure selectable from 25 psig to available
constant Line Pressures.
2. Inspiratory Flowrate Selectable with Programmable acceleration, from OFF to over
125 liters per minute with 40 psig Operational Pressure.
3. Demand CPAP (continuous positive airway pressure) during Spontaneous Breathing
from OFF to 20 cm H2O. (Optional to over 50 cm H20).
4. Low Rate Convective Cycling from 1 to over 40 cycles/minute with Inspiratory/Expiratory
Intervals (I/E ratios) programmable from positive to negative Ratios within acceptable
Clinical Parameters.
5. High Rate Diffusive Cycling from 200 to over 800 cycles/minute with i/e ratios from 1:1 to 1:3.
6. Three stage Sinusoidal Oscillatory Equilibrium Scheduling, for Pulmonary Conformance.
7. Direct Reading lower rate Proximal Airway Pressure presentation with higher frequency
presentation dampened. Selectable Integrated manometer for Ventilatory
Pressure Averaging.
8. Nebulizer flow is adjustable from OFF to Line Pressure for Aerosol Therapy etc.
9. Selectable integrated high pressure Failsafe Sensitivity with Aural Alerting and
Manual Reset.
10. Manual Push Button activation of Inspiratory Flowrate, for selectable Volume
Delivery Override.
11. Full Ambient flow clutching from 25 to over 125 cm H20 as Governed by Operational
Pressure Selection.
12. Proximal Airway delivery pressure range, from 1 to over 125 cm H20.
13. Cyclic Lung Volume Exchange potentials, from 5 cc to over 10 liters per programmed
Inspiratory Interval.
14. Delivered Volume is equivalent to Oscillatory Flow multiplied by time modified by
(anti-barotrauma) ambient referenced fluidic clutching provided by the Phasitron
Injector/Exhalation Valve Venturi.
15. Phasitron VDR® Breathing Circuit with (near zero compliance) Proximal Airway
Flow Generation.
16. Self Contained high flow Oxygen/Air Blending for selectable FI02.
17. Simplex Control Knob System for Program Initiation, Charting and Documentation.
18. Programmable accumulative Sub Tidal Volume Delivery to Oscillatory Equilibrium.
19. Controlled Percussive Pulse Generation for increasing the Intrapulmonary Diffusive
Component of Ventilation during the progressive increase in Lung Volumes, to Oscillatory
Equilibrium, followed by the passive exhalation of a traditional Tidal volume; down to the
programmed static or oscillatory baseline.
20. Pulse Frequency Impulse Force is controlled by Pulsatile Flowrate.
21. Accessory Monitron® Proximal Airway Wave Form Analyzer with Low and High
Cycling Frequencies as well as Pressure Rise/Drop monitoring of the Proximal
Airway Pressures. |